New Features and Enhancements

System Tray Icon Prefetch Status

Prefetch status' have been added to the viewer's system tray icon.
NOTE: Available only for eVolution viewers used with version 7 servers.

Prefetch Status Status Description
Viewer actively polling for studies to download from configured prefetch lists .
Viewer actively downloading studies from configured prefetch lists.
Viewer has completed downloading of all content on configured prefetch lists.
Viewer has encountered an error while prefetching.  Use system tray context menu to view error details.

 

Resolved Issues

Incorrect Spine Label Placement

The frame of reference refactoring needed to support 3D linear measurements was not applied to spine label's coordinate system, causing the placement of spine labels in related images to appear outside the image.  This has been resolved.

Incorrect Application of Spatial Registration Objects

A miscalculation when linking deformable registration objects resulted in a buffer overread and placing the intersecting points beyond the end of the series.

Cache Information Did Not Contain Accession or Patient ID

The accession number and PID values were not stored in the cache map, making them unavailable to WMI queries after viewer restarts.  This has been resolved.
NOTE: Applies only for eVolution viewers used with version 7 servers.

Viewer Crash: Loading Hanging Protocol with Image Enhancement

If a hanging protocol specifies an enhanced image and the viewer has not created the enhanced image when the viewer starts processing the HP, the viewer could crash.  This has been resolved.

Spatial Registration Results Not Auto Linking When Localizer Images Exist

If a series contains images in multiple orientations, meaning the series is a collection of views or has a localizer image preceding serial images, auto-linking wasn't always applied when loading a series into an image frame.  This has been resolved.

Viewer Crash: Deleted Images

Deleting reconstructed images (exported from a plugin frame) from the thumbnail panel while the series appears in a visible image frame failed to clear the frame, resulting in a crash when selecting the image frame.  This has been resolved.

Mammography Skin Line Detection Could Fail

When the skin line detection software encountered burned-in markers, the algorithm could fail.  This has been resolved.