New Features and Enhancements

Viewer Build 34

NA

Viewer Build 33

NA

Viewer Build 32

Show annotation values setting expanded

The Options/Annotations/Modality-dependent settings page now includes separate "Show value" settings for the following annotation tools: measure distance, measure angle, Hounsfield, ROI, and Free ROI.  The default value for all the new settings is enabled, except ROI tools for MG modality

Resolved Issues

Viewer Build 34

Prevent split mammo multiframe objects into individual series

Multiframe objects are included in the image count when checking the split series length, but are treated as a single image object when splitting, meaning multiframe objects will not be separated into individual series in the thumbnail panel.

Left side thumbnail panel scrolling incorrect for multi monitor configuration

When the thumbnail panel is docked along the left side of the window on a multiple monitor workstation, scrolling the thumbnail panel would eventually change from a vertical scroll to a horizontal scroll, corrupting the screen layout.

Report tabs aren't properly updated after opening multiple studies

When opening a viewer session with multiple studies, at least one of which is in the Final state (ie, a prior study), the primary study's report panel's tabs could be mislabeled until the user manually selects one of the tabs.  This has been resolved.

Viewer shows connection lost message when the loading was interrupted

If the viewer session closes before the study pump releases all the thumbnail references, the service component could crash resulting in a lost connection message and no data downloads.  This has been resolved.

Scanning application doesn't close the dialog after uploading

When the scanning operation completed, the scanning application panel was not closed automatically.  This has been resolved.

Viewer crash: Reference spatial registration is crashing

Spatial registration objects containing references to three or more frames of reference were linked incorrectly, resulting to a viewer crash applied.  This has been resolved.

Duplicated CAD objects

When appending studies with CAD objects to an active viewer session, the viewer appended CAD objects from all studies to the thumbnail panel, duplicating those that already existed.  This has been resolved.

Viewer Build 33

Mammography view progression mode gets out of sync on linked images

Mammography images linked using Smart Link could get out of sync when using the progression tool, especially if one of the linked images failed to identify the anatomical structure in the image.  This has been resolved.

Viewer crashes with annotation flip/rotation

Altering a linear or angular measurement on an image whose orientation has been modified from the default (by the auto-orientation feature) could cause the measurement's text box dimensions to become negative, resulting in lines left on the screen and possibly a viewer crash.  This has been resolved.

Viewer crashes with certain name format configurations

Using a name component (eg, Family Name) multiple times in the name format definition could result in an overrun buffer, causing the viewer to crash.  This has been resolved.

Invert anatomy or show skin line crashes viewer

Inverting the greyscale on a mammography image before it completely decompresses initiated recursive calls to the skin line detection function, causing a stack overflow and a viewer crash.  This has been resolved.

Viewer Build 32

DeepHealth CAD border calculation error

Two CAD finding ROIs that shared a common edge would appear overlapped due to a calculation error when drawing the double-line ROI border. This has been resolved.

Cursor over an annotation stops mouse wheel scrolling

If the mouse cursor was positioned directly over an annotation ROI or line graphic, the mouse wheel scroll feature would not activate. This has been resolved.

Viewer Build 31

Spine Label Plane Settings Not Applied

The spine label plane setting displayed under the Settings tab on the popup panel was not recognized, resulting in the inability to enable or disable the feature when applying spine labels.

Center Point Visible on Oval SR Objects

After adding center points to ROI annotations, the center point was incorrectly applied to mammo CAD ROI annotations. This has been resolved.

UI Controls for Saving Presentation States at Closing Not Working

When closing a viewer session containing unsent presentation states intended to be uploaded as secondary capture objects, the close panel includes the option to send them. This setting was not active, meaning changes made in the close panel were ignored.  This has been resolved.

Incorrect Sort Order When Series is Augmented as Axial

When a series is auto-oriented in the axial direction, the order direction could be misassigned,resulting in the wrong sort order.  This has been resolved.

Scrolling Lag Over Image Frame Borders