PACS
Desktop Viewer
The Options/Annotations/Modality-dependent settings page now includes separate "Show value" settings for the following annotation tools: measure distance, measure angle, Hounsfield, ROI, and Free ROI. The default value for all the new settings is enabled, except ROI tools for MG modality
Two CAD finding ROIs that shared a common edge would appear overlapped due to a calculation error when drawing the double-line ROI border. This has been resolved.
If the mouse cursor was positioned directly over an annotation ROI or line graphic, the mouse wheel scroll feature would not activate. This has been resolved.
The spine label plane setting displayed under the Settings tab on the popup panel was not recognized, resulting in the inability to enable or disable the feature when applying spine labels.
After adding center points to ROI annotations, the center point was incorrectly applied to mammo CAD ROI annotations. This has been resolved.
When closing a viewer session containing unsent presentation states intended to be uploaded as secondary capture objects, the close panel includes the option to send them. This setting was not active, meaning changes made in the close panel were ignored. This has been resolved.
When a series is auto-oriented in the axial direction, the order direction could be misassigned,resulting in the wrong sort order. This has been resolved.
If the mouse hovers over the frame border while scrolling via the left-mouse drag tool, scrolling would stop. This has been resolved.